Medical Device Recall: MiniMed 780G Insulin Pump, REF: MMT-1885, MMT-1886, MMT-1895, MMT-1896
Medtronic MiniMed, Inc. · May 27, 2026
Reason for Recall
A software anomaly can occur when the pump is paired with the Instinct sensor. Under certain circumstances, the pump unexpectedly stops delivering insulin and displays two error messages back-to-back: Pump Error 53 followed immediately by Pump Error 23. Insulin delivery will stop which may lead to delayed therapy and an increased risk of hyperglycemia or diabetic ketoacidosis (DKA).
Distribution
International distribution to the countries of Spain, Great Britain, Netherlands, Italy.
Quantity Affected
778
Source: FDA (openFDA Device Enforcement)
Recall number: Z-2698-2026