Class II

Medical Device Recall: Medtronic SynchroMed Flex Infusion Mode - A810 Clinician Programmer Software Application, loaded on the following devices: 1. PROG CT900A CLINICIAN TABLET US, Model Number CT900A; 2. PROG CT900B

Medtronic Neuromodulation · April 8, 2026

Reason for Recall

Software issue: When programming the SynchroMed Pump with the A810 CP App in Flex Infusion Mode, the infusion steps should automatically be arranged in consecutive order, from earliest to latest. However, in rare cases, the steps may be displayed out of order. If this occurs, the programmed Flex Infusion schedule will not be delivered at the intended time, if at all.

Product Description

Medtronic SynchroMed Flex Infusion Mode - A810 Clinician Programmer Software Application, loaded on the following devices: 1. PROG CT900A CLINICIAN TABLET US, Model Number CT900A; 2. PROG CT900B CLINICIAN TABLET EU -UK, Model Number CT900B; 3. PROG CT900C CLINICIAN TABLET UK, Model Number CT900C; 4. PROG CT900D CLINICIAN TABLET GLOBAL, Model Number CT900D; 5. PROG CT900E CLINICIAN TABLET GLBL SMSNG, Model Number CT900E; 6. PROG CT900F CLINICIAN TABLET GLBL SMSNG, Model Number CT900F; implanted programmable infusion pump

Distribution

Worldwide distribution.

Quantity Affected

10165 units

Source: FDA (openFDA Device Enforcement)

Recall number: Z-2201-2026

Status: ongoing

SafeCheck provides drug interaction information from FDA-approved product labeling (openFDA). This is for informational purposes only and is not medical advice. Always consult your healthcare provider before making medication decisions.