Class II

Medical Device Recall: MC3 VitalFlow Console, REF 58100; Blood pump of ecmo

Medtronic Perfusion Systems · October 3, 2025

Reason for Recall

As of August 6, 2025, Medtronic has received eleven reports of VitalFlow Consoles displaying an E70 error code during normal operation. When this occurs, the touch screen may become temporarily unresponsive and go blank for up to two minutes before recovering to full functionality. Importantly, the console continues to maintain set pump speed and function throughout.

Distribution

US Nationwide distribution.

States Affected

NATIONWIDE

Quantity Affected

167 units

Source: FDA (openFDA Device Enforcement)

Recall number: Z-0509-2026

Status: ongoing