Medical Device Recall: VERICIS, Merge Cardio, Model/Catalog Number: Version 10.x, Software Version: 10.0, 10.1.1.1, 10.1.2, 10.2.0, 10.2 P2 , 10.3; The VERICIS is an integrated cardiovascular information system classified a
Merge Healthcare, Inc. · February 19, 2025
Reason for Recall
When following a specific less typical workflow in which the user remeasures using the ultrasound device or manually edits primitive or derived measurement values directly in Clinical Reporting, derived measurements may be inconsistent with the associated primitive measurements in the final patient report.
Product Description
VERICIS, Merge Cardio, Model/Catalog Number: Version 10.x, Software Version: 10.0, 10.1.1.1, 10.1.2, 10.2.0, 10.2 P2 , 10.3; The VERICIS is an integrated cardiovascular information system classified as a picture archiving and communications system.
Distribution
US
Quantity Affected
11
Source: FDA (openFDA Device Enforcement)
Recall number: Z-1399-2025
Status: ongoing