Class II

Medical Device Recall: The D*Clot¿ HD Rotational Thrombectomy System is a percutaneous catheter-based system available in 6 French. The disposable system consists of an outer sheath through which a rotatable radiopaque spi

Mermaid Medical A/S · May 1, 2024

Reason for Recall

Internal product testing failed bubble leak testing (ASTM F2096), which indicates the sterility barrier may not be fully intact.

Product Description

The D*Clot¿ HD Rotational Thrombectomy System is a percutaneous catheter-based system available in 6 French. The disposable system consists of an outer sheath through which a rotatable radiopaque spiral shaft extends ending in an eccentric blunt distal tip. Once activated, the shaft rotates at approximately 10.000 RPMs to break thrombus into smaller particles, which can be simultaneously aspirated using a dedicated vacuum syringe. The catheter is a stand-alone system and is powered by a 4.5 VDC alkaline battery pack in the control unit.

Distribution

Worldwide - US Nationwide distribution in the states of OH, MO, CO, NC, TX, MA, GA, FL, IL, IN, MD, PA, VA, and the country of Chile.

States Affected

NATIONWIDE

Quantity Affected

190 devices

Source: FDA (openFDA Device Enforcement)

Recall number: Z-2213-2024

Status: ongoing

SafeCheck provides drug interaction information from FDA-approved product labeling (openFDA). This is for informational purposes only and is not medical advice. Always consult your healthcare provider before making medication decisions.