Medical Device Recall: Medtronic Pipeline Vantage Embolization Device with Shield Technology, REF: PED3-027-350-12, PED3-027-350-14, PED3-027-350-16, PED3-027-350-20, PED3-027-350-25, PED3-027-400-12, PED3-027-400-14, PED3-
Micro Therapeutics, Inc. · January 30, 2025
Reason for Recall
Use of embolization device with diameters greater-than-or-equal-to 4mm may lead to incomplete wall apposition and/or braid deformation (fish-mouthing/braid narrowing/braid collapse) noted during procedure and post-procedure, typically noted at 6-12 month imaging follow-up, potentially can lead to thrombosis and/or serious adverse events. Higher risk in females less-than-or-equal-to 45 years of age
Product Description
Medtronic Pipeline Vantage Embolization Device with Shield Technology, REF: PED3-027-350-12, PED3-027-350-14, PED3-027-350-16, PED3-027-350-20, PED3-027-350-25, PED3-027-400-12, PED3-027-400-14, PED3-027-400-16, PED3-027-400-18, PED3-027-400-20, PED3-027-400-25, PED3-027-400-30, PED3-027-450-12, PED3-027-450-14, PED3-027-450-16, PED3-027-450-18, PED3-027-450-20, PED3-027-450-25, PED3-027-450-30, PED3-027-450-40, PED3-027-500-14, PED3-027-500-16, PED3-027-500-18, PED3-027-500-20, PED3-027-500-25, PED3-027-500-30, PED3-027-500-40, PED3-027-550-16, PED3-027-550-18, PED3-027-550-20, PED3-027-550-30, PED3-027-550-40, PED3-027-550-50, PED3-027-600-16, PED3-027-600-18, PED3-027-600-20, PED3-027-600-30, PED3-027-600-40, PED3-027-600-50
Distribution
Worldwide distribution: US (nationwide): PR, OK, CA, MD, FL, NY, WA, MI, AL, NV, MN, NJ, IL, OH, AK, CT, TN, MO, TX, ID, AZ, WI, PA, OR, SD, VA, CO, GA, KY, IN, AR, MA, LA, NE, MT, ME, SC, DC, NC, ND, NM, IA, UT, KS, WV, NH, HI, RI. And OUS (foreign) countries of: Algeria, Argentina, Armenia, Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Chile, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Jordan, Kazakhstan, Kuwait, Latvia, Lebanon, Lithuania, Luxembourg, Malaysia, Martinique, Mexico, Republic Of Moldova, Montenegro, Morocco, Netherlands, New Zealand, North Macedonia, Northern Ireland, Norway, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Vietnam
States Affected
NATIONWIDE
Quantity Affected
18,245
Source: FDA (openFDA Device Enforcement)
Recall number: Z-1245-2025
Status: ongoing
SafeCheck provides drug interaction information from FDA-approved product labeling (openFDA). This is for informational purposes only and is not medical advice. Always consult your healthcare provider before making medication decisions.