Class II

Medical Device Recall: BOBBY Balloon Guide Catheter REF: BOB-895-CN BOB-895-EMEA BOB-895-EMEA-AL BOB-895-US MV-BGA01095 For use in facilitating the insertion and guidance of an intravascular catheter into a selecte

MICROVENTION INC. · April 30, 2026

Reason for Recall

Due to an increase rate of balloon pinhole defects.

Product Description

BOBBY Balloon Guide Catheter REF: BOB-895-CN BOB-895-EMEA BOB-895-EMEA-AL BOB-895-US MV-BGA01095 For use in facilitating the insertion and guidance of an intravascular catheter into a selected blood vessel in the peripheral and neurovascular systems.

Distribution

Worldwide - U.S. Nationwide distribution in the states of AL, AR, AZ, CA, FL, GA, IA, IL, KS, KY, LA, MI, MN, MO, MT, NC, NH, NM, NY, OH, PA, SD, TN, TX, WA, and WI. The countries of Europe - Austria, Belgium, Denmark, Finland, France, Germany, Greece, Iceland, Ireland, Italy, Lithuania, Netherlands, Norway, Poland, Romania, Slovenia, South Africa, Spain, Sweden, Switzerland, Ukraine, and United Kingdom. Asia - China, Japan, Jordan, New Zealand, South Korea, and United Arab Emirates.

States Affected

NATIONWIDE

Quantity Affected

5500 devices

Source: FDA (openFDA Device Enforcement)

Recall number: Z-2912-2026