Class I
Medical Device Recall: Multi Joystick R-net. Electrical wheelchair component.
mo-Vis BVBA · July 10, 2025
Reason for Recall
Devices with firmware versions 2.3 and lower may experience timing variations between software tasks, which could result in skipping the necessary neutral check and the wheelchair producing unintended movement upon power-up when the joystick is out of neutral.
Distribution
US Nationwide distribution in the state of TX.
States Affected
NATIONWIDE
Quantity Affected
3 units
Source: FDA (openFDA Device Enforcement)