Medical Device Recall: Handy Solutions Neck & Shoulders Heating Pad, Number: 25607, Manufactured by Ninbo Unico Products Co., Ltd
Navajo Manufacturing Company · February 18, 2026
Reason for Recall
If heating pad is folded while in use, the power density increases in the folded area causing excessive temperatures, which can also be caused by placing pad under the body, then the high limit device is unable to sense the higher temperature so power is not terminated. May cause a burning smell prior to melting/burning potentially causing burns to persons and property.
Distribution
US Nationwide distribution in the states of CA, MS, FL, IN, LA, VA, TX.
States Affected
NATIONWIDE
Quantity Affected
28,457
Source: FDA (openFDA Device Enforcement)
Recall number: Z-1588-2026
Status: ongoing