Class I

Medical Device Recall: Handy Solutions Neck & Shoulders Heating Pad, Number: 25607, Manufactured by Ninbo Unico Products Co., Ltd

Navajo Manufacturing Company · February 18, 2026

Reason for Recall

If heating pad is folded while in use, the power density increases in the folded area causing excessive temperatures, which can also be caused by placing pad under the body, then the high limit device is unable to sense the higher temperature so power is not terminated. May cause a burning smell prior to melting/burning potentially causing burns to persons and property.

Distribution

US Nationwide distribution in the states of CA, MS, FL, IN, LA, VA, TX.

States Affected

NATIONWIDE

Quantity Affected

28,457

Source: FDA (openFDA Device Enforcement)

Recall number: Z-1588-2026