Class II

Medical Device Recall: NeuroSync EYE-SYNC containing Pico Neo 2 Headsets. Each unit consists of a Pico Neo 2 headset with embedded sensors, a paired Android tablet, and proprietary software, packaged in a portable field cas

NeuroSync, Inc. · September 19, 2023

Reason for Recall

Eye syncing issues causing the device to be inoperable. Unable to be repaired due to discontinuation of support for the Pico 2 platform, which was no longer in production or serviceable as of August 31, 2023.

Product Description

NeuroSync EYE-SYNC containing Pico Neo 2 Headsets. Each unit consists of a Pico Neo 2 headset with embedded sensors, a paired Android tablet, and proprietary software, packaged in a portable field case. Eye-tracking headset system is intended for recording, viewing, and analyzing eye movements in support of identifying visual tracking impairment in human subjects. The EYE-SYNC is intended to record, measure, and analyze eye movements as an aid in the diagnosis of concussion, also known as mild traumatic brain injury (mTBI).

Distribution

United States Nationwide distribution in the states of New York (NY), Texas (TX), and Georgia (GA).

States Affected

NATIONWIDE

Quantity Affected

27 units

Source: FDA (openFDA Device Enforcement)

Recall number: Z-2114-2025

Status: ongoing