Class II

Medical Device Recall: Integra Miltex CryoSolutions Cartridges 10 Pack (23.5g N2O each), Model: C-CA-23, Item: 33516;

New Medical Technologies Gmbh · December 19, 2024

Reason for Recall

During the assembly of the cartridge to its connecting body part, the valve may open prematurely resulting in partial or complete emptying of the cartridge before its intended use.

Distribution

Worldwide - US Nationwide distribution in the states of AR, CA, CO, FL, GA, IL, IN, KY, LA, MA, MD, MN, MO, NC, NE, NJ, NV, NY, OH, PA, SC, TN, TX, VA, WA and the country of Canada.

States Affected

NATIONWIDE

Quantity Affected

430 units (43 packs of 10)

Source: FDA (openFDA Device Enforcement)

Recall number: Z-1103-2025

Status: ongoing

SafeCheck provides drug interaction information from FDA-approved product labeling (openFDA). This is for informational purposes only and is not medical advice. Always consult your healthcare provider before making medication decisions.