Class II

Medical Device Recall: Galaxy bronchoscope (Product GALB-001) used with Instructions for use, Part: 10000752, part of the Noah Medical Galaxy System (GAL-001)

Noah Medical · October 25, 2024

Reason for Recall

Not all biopsy tools meant to fit bronchoscope channel (2.1mm inner diameter/length greater or equal to 1100mm) will pass through if scope channel positioned in sharp bend (approx. greater than 100-deg.), excessive force may result in tissue injury/pneumothorax; instruments with long rigid distal sections may exhibit insertion resistance when bronchoscope bending section significantly articulated.

Distribution

US Nationwide Distribution: SC, OH, PA, TN, MO, IN, CA, WV, NY, MD, MA, NJ, ID, CT, NC

States Affected

NATIONWIDE

Quantity Affected

26

Source: FDA (openFDA Device Enforcement)

Recall number: Z-0679-2025

Status: ongoing