Class I

Medical Device Recall: NOxBOXi Nitric Oxide Delivery System; Model Numbers: NOXBOX-I

NOXBOX LTD · September 9, 2025

Reason for Recall

Fluctuations may be more likely to occur if the total flow through the device sensor (used by the device to determine flow) is less than the minimum device requirement of 0.5 LPM. In addition, dose fluctuations have been observed when Bunnel LifePulse HFJV system or conventional ventilators have the following device settings: low service pressures generally less than 3.0 PSI, a pressure difference greater than 5.0cm H2O; or at breath per minute rates of 300 (5 Hz).

Distribution

Domestic: TN

States Affected

TN

Quantity Affected

1667 units

Source: FDA (openFDA Device Enforcement)

Recall number: Z-0028-2026

Status: ongoing

SafeCheck provides drug interaction information from FDA-approved product labeling (openFDA). This is for informational purposes only and is not medical advice. Always consult your healthcare provider before making medication decisions.