Class II

Medical Device Recall: NxStage Cartridge Express with Speedswap, Catalog No. CAR-535. An accessory device for facilitating the removal and replacement of a flow-compromised filter during treatment of acute and chronic re

NxStage MDS Corporation · January 29, 2024

Reason for Recall

Product is recalled due to reports of blood leaking or spraying from the arterial line on the cartridge.If this failure occurs, a patient may experience uncontrolled blood loss or a healthcare professional could be exposed to infectious patient blood.

Product Description

NxStage Cartridge Express with Speedswap, Catalog No. CAR-535. An accessory device for facilitating the removal and replacement of a flow-compromised filter during treatment of acute and chronic renal failure or fluid overload.

Distribution

US domestic distribution to Dialysis Centers located in AZ, CA, CO, GA, IL, IN, MA, MD, MI, MN, MO, MS, NC, NJ, NY, TN, TX, VA, WI. No international distribution.

States Affected

AZ, CA, CO, GA, IL, IN, MA, MD, MI, MN, MO, MS, NC, NJ, NY, TN, TX, VA, WI

Quantity Affected

693

Source: FDA (openFDA Device Enforcement)

Recall number: Z-1191-2024

Status: ongoing