Class II

Medical Device Recall: NxStage PureFlow-B Solution, Premixed Dialysate for Hemodialysis, Device Catalog Numbers RFP-400, RFP-401, RFP-402, RFP-403, RFP-404, RFP-405, RFP-406, RFP-453, RFP-454, RFP-456

NxStage Medical Inc · May 14, 2024

Reason for Recall

The smaller chamber of the two chamber bag is bursting when nurses are attempting to mix the two chambers prior to use. The electrolyte fluid in the small chamber of the NxStage PureFlow B Solution has a pH of less than 2.0. The solution can be harmful if it comes in contact with the eyes or skin, and/or as a result of a slip or fall injury due to spilled fluid.

Distribution

US Nationwide distribution.

States Affected

NATIONWIDE

Quantity Affected

1316936 units

Source: FDA (openFDA Device Enforcement)

Recall number: Z-2160-2024

Status: ongoing