Class II

Medical Device Recall: Brand Name: EVIS EXERA III Gastrointestinal Videoscope GIF-1TH190 Product Name: Olympus GIF-1TH190 Model/Catalog Number: GIF-1TH190 Software Version: N/A Product Description: This instrument is in

Olympus Corporation of the Americas · September 18, 2024

Reason for Recall

To better define the drying procedures described in the IFU to improve the instructions for easier understanding of each drying step after human errors and concerns were identified when completing a Human Factors evaluation of the GIF-1TH190.

Product Description

Brand Name: EVIS EXERA III Gastrointestinal Videoscope GIF-1TH190 Product Name: Olympus GIF-1TH190 Model/Catalog Number: GIF-1TH190 Software Version: N/A Product Description: This instrument is intended to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories (such as biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery.

Distribution

Domestic: US Nationwide Distribution.

States Affected

NATIONWIDE

Quantity Affected

1749 units

Source: FDA (openFDA Device Enforcement)

Recall number: Z-0498-2025

Status: ongoing