Medical Device Recall: Brand Name: Olympus Everest Bipolar Cutting Forceps w/Cord, 5mm 33 cm Product Name: Everest Bipolar Cutting Forceps Model/Catalog Number: 3006 Product Description: The OLYMPUS Everest Bipolar Cutti
Olympus Corporation of the Americas · February 25, 2026
Reason for Recall
Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use.
Product Description
Brand Name: Olympus Everest Bipolar Cutting Forceps w/Cord, 5mm 33 cm Product Name: Everest Bipolar Cutting Forceps Model/Catalog Number: 3006 Product Description: The OLYMPUS Everest Bipolar Cutting Forceps are intended to be passed through a 5 mm cannula. Coagulation is achieved using electrosurgical energy under visualization. The device is intended to be used with the bipolar outputs of compatible generators.
Distribution
Worldwide - US Nationwide and the countries of Canada, Mexico, EU, Australia, China, Singapore, Korea, Japan.
States Affected
NATIONWIDE
Quantity Affected
164 units
Source: FDA (openFDA Device Enforcement)
Recall number: Z-1639-2026