Class II

Medical Device Recall: Electrosurgical, Cutting and Coagulation and Accessories. Electrosurgical Generator ESG-410. Model Number: WA91307C. The ESG-410 Electrosurgical Generator is an electrosurgical generator which uti

Olympus Corporation of the Americas · July 25, 2025

Reason for Recall

Potential for the ESG-410 Electrosurgical Generator to display an error message: "E0662 System Error", followed by either automatic single reboots of the device, or continuous reboot 'loops'.

Product Description

Electrosurgical, Cutting and Coagulation and Accessories. Electrosurgical Generator ESG-410. Model Number: WA91307C. The ESG-410 Electrosurgical Generator is an electrosurgical generator which utilizes monopolar and bipolar high frequency current and supports ultrasonic instruments during open surgery, laparoscopic surgery, including single-site surgery, and endoscopic surgery.

Distribution

US Nationwide distribution.

States Affected

NATIONWIDE

Quantity Affected

434 units

Source: FDA (openFDA Device Enforcement)

Recall number: Z-2339-2025

Status: ongoing

SafeCheck provides drug interaction information from FDA-approved product labeling (openFDA). This is for informational purposes only and is not medical advice. Always consult your healthcare provider before making medication decisions.