Medical Device Recall: Electrosurgical, Cutting and Coagulation and Accessories. Electrosurgical Generator ESG-410. Model Number: WA91307C. The ESG-410 Electrosurgical Generator is an electrosurgical generator which uti
Olympus Corporation of the Americas · July 25, 2025
Reason for Recall
Potential for the ESG-410 Electrosurgical Generator to display an error message: "E0662 System Error", followed by either automatic single reboots of the device, or continuous reboot 'loops'.
Product Description
Electrosurgical, Cutting and Coagulation and Accessories. Electrosurgical Generator ESG-410. Model Number: WA91307C. The ESG-410 Electrosurgical Generator is an electrosurgical generator which utilizes monopolar and bipolar high frequency current and supports ultrasonic instruments during open surgery, laparoscopic surgery, including single-site surgery, and endoscopic surgery.
Distribution
US Nationwide distribution.
States Affected
NATIONWIDE
Quantity Affected
434 units
Source: FDA (openFDA Device Enforcement)
Recall number: Z-2339-2025
Status: ongoing