Medical Device Recall: EZDilate Fixed Wire Balloon 13.5-14.5-15.5. Indicated for endoscopic dilation of strictures of the esophagus in adults and adolescents (>12 years). Model: BD-400P-1580
Olympus Corporation of the Americas · November 10, 2023
Reason for Recall
Product balloon issues: inflating, deflating and/or retrieving the devices; reports of bursting and leaking and events involving foreign body in patient and prolonged procedure.
Distribution
Nationwide. Foreign: AU BR BR CA DE HK IN JP SG
States Affected
NATIONWIDE
Quantity Affected
5232 units
Source: FDA (openFDA Device Enforcement)
Recall number: Z-0936-2024
Status: ongoing