Class II

Medical Device Recall: Olympus BiCOAG Hemostasis Probe 7Fr Coaxial Connector. Model/Catalog Number: CD-B620LA.

Olympus Corporation of the Americas · August 19, 2026

Reason for Recall

Potential for detachment of ceramic tip due to manufacturing without the designated amount of adhesive.

Distribution

Worldwide distribution - US Nationwide and the countries of Canada, Latin America, Singapore, Hong Kong, and EU.

States Affected

NATIONWIDE

Quantity Affected

2,142 units

Source: FDA (openFDA Device Enforcement)

Recall number: Z-3246-2026

Status: ongoing

SafeCheck provides drug interaction information from FDA-approved product labeling (openFDA). This is for informational purposes only and is not medical advice. Always consult your healthcare provider before making medication decisions.