Class II

Medical Device Recall: IVX Station Reconstitution Transfer Set. Model Number: 298921215. Packaging: 1 set per White Tyvek Pouch, 5 Tyvek Pouches per shelf box, 10 shelf boxes per case box. IVX Station Reconstitution Tran

Omnicell, Inc. · February 20, 2025

Reason for Recall

IVX Station Reconstitution Transfer Sets were shipped in advance of the IVX Station upgrade and not intended for clinical use but are engineering devices intended for use during set up and system calibration. The items were not shipped in a validated packing configuration and sterility cannot be ensured.

Product Description

IVX Station Reconstitution Transfer Set. Model Number: 298921215. Packaging: 1 set per White Tyvek Pouch, 5 Tyvek Pouches per shelf box, 10 shelf boxes per case box. IVX Station Reconstitution Transfer Set is used for the reconstitution of drug vials in hospital pharmacies. It is indicated for use with the IVX Station Pharmacy Compounding System.

Distribution

Distributed in the US to Alabama

States Affected

IN

Quantity Affected

3

Source: FDA (openFDA Device Enforcement)

Recall number: Z-1532-2025

Status: ongoing

SafeCheck provides drug interaction information from FDA-approved product labeling (openFDA). This is for informational purposes only and is not medical advice. Always consult your healthcare provider before making medication decisions.