Class II

Medical Device Recall: OrganOx metra -Intended as a Normothermic Machine Perfusion system of donor transport livers. Model/Catalog Number: D0003 The OrganOx metra is a transportable device intended to be used to susta

OrganOx Ltd · October 15, 2024

Reason for Recall

Battery movement led to damage of one of the battery cells causing it to short to the battery retention bracket

Product Description

OrganOx metra -Intended as a Normothermic Machine Perfusion system of donor transport livers. Model/Catalog Number: D0003 The OrganOx metra is a transportable device intended to be used to sustain donor livers destined for transplantation in a functioning state for a total preservation time of up to 12 hours.

Distribution

California, Texas, Ohio, Florida, New York, Massachusetts, Virginia, Oklahoma, DC, South Carolina and Georgia.

States Affected

DC

Quantity Affected

20 units

Source: FDA (openFDA Device Enforcement)

Recall number: Z-0501-2025

Status: ongoing