Medical Device Recall: VITROS Immunodiagnostic Products Vitamin B12 Range Verifier - In vitro diagnostic use ly. For use in verifying the calibration range of the VITROS ECi/ECiQ/3600 Immunodiagnostic Systems and the VITROS
Ortho-Clinical Diagnostics, Inc. · May 8, 2024
Reason for Recall
The results generated by the VITROS Immunodiagnostic Products Vitamin B12 Range Verifier Lot# 1200 RV Level 2 may cause range verification failure. The issue is detectable by the user when the product fails to meet the required range during processing of RV Level 2 of this lot. There is a potential for a delay in patient test results should the user have a need to switch to an alternate method of range verification including use of an alternate lot. As this issue affects the upper limit range verification, there is no impact to VITROS Immunodiagnostic Products Vitamin B12 assay performance. Were the assay to not perform as expected, the failure would be detectable by Quality Control. Past patient sample results are unaffected, and no review of reported results is required.
Product Description
VITROS Immunodiagnostic Products Vitamin B12 Range Verifier - In vitro diagnostic use ly. For use in verifying the calibration range of the VITROS ECi/ECiQ/3600 Immunodiagnostic Systems and the VITROS 5600/XT 7600 Integrated Systems when used for the quantitative measurement of vitamin B12 - Product Code: 6800872
Distribution
Worldwide - US Nationwide and the country of Columbia.
States Affected
NATIONWIDE
Quantity Affected
76 units
Source: FDA (openFDA Device Enforcement)
Recall number: Z-2019-2024
Status: ongoing