Medical Device Recall: VITROS XT 3400 Chemistry System, Catalog No. 6844458
Ortho-Clinical Diagnostics, Inc. · January 26, 2024
Reason for Recall
A software defect affecting VITROS Systems running VITROS Software Versions 3.8.1 causes QC baseline statistics to not update as expected when changed by the user. This may cause erroneous patient results to be reported.
Distribution
Worldwide - US Nationwide distribution in the states of AZ, CA, CT, FL, IA, ID, IL, IN, KY, LA, MO, NE, NY, OH, OR, SC, SD, TX, WA, WI, WV and the countries of Australia, Brazil, Canada, China, Colombia, India, Mexico, Singapore, United Kingdom, Italy, Spain, Russia, France, and Belgium.
States Affected
NATIONWIDE
Quantity Affected
7 US; 4 OUS
Source: FDA (openFDA Device Enforcement)
Recall number: Z-1227-2024
Status: ongoing