Medical Device Recall: VITROS Performance Verifier II

Ortho-Clinical Diagnostics · February 27, 2024

Reason for Recall

The issue was mismatched diluent lots assembled into VITRO Chemistry Products Performance Verifier I and II. This issue has the potential to cause erroneous Quality Control (QC) results that could lead to testing delays and erroneous patient results. This issue caused positive or negative bias in the quantification of sodium (Na+) and total carbon dioxide (ECO2). The incorrect diluent into the Performance Verifier II exhibited a potential negative bias of up to 6 mmol/L for ECO2 and Na+ testing.

Distribution

Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway, Portugal, Russia, Singapore, Spain, Sweden, The Netherlands, United Kingdom.

States Affected

NATIONWIDE

Quantity Affected

11,224 units

Source: FDA (openFDA Device Enforcement)

Recall number: Z-1373-2024

Status: ongoing