Class II

Medical Device Recall: Orthopantomograph OP300 dental computed tomography x-ray system. The X-ray device produces conventional 2D X-ray images and Xray projection images for the reconstruction of a 3D view. The device i

Palodex Group Oy · October 28, 2021

Reason for Recall

It was noticed in production on 28.10.2021, that the final testing database gives a pass result for Pan ADC repeatability test even when the test result is not within the set acceptance criteria (repeatability within 4%).

Product Description

Orthopantomograph OP300 dental computed tomography x-ray system. The X-ray device produces conventional 2D X-ray images and Xray projection images for the reconstruction of a 3D view. The device is operated and used by qualified healthcare professionals.

Distribution

US Nationwide-Worldwide

States Affected

NATIONWIDE

Quantity Affected

103

Source: FDA (openFDA Device Enforcement)

Recall number: Z-2413-2024

Status: ongoing