Class II

Medical Device Recall: M5 Corpus Power Wheelchair REF,120032

Permobil Inc · June 19, 2025

Reason for Recall

This recall is being reported retroactively. Power Platform firmware is missing specific drive inhibit logic. This issue may allow the chair to drive under restricted conditions, and may create an unstable condition in certain circumstances.

Distribution

Worldwide - US: Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KY, ME, MI, MN, MO, MS, MT, NC, ND, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI, WV, WY and the country of CANADA.

States Affected

NATIONWIDE

Quantity Affected

715 wheelchairs

Source: FDA (openFDA Device Enforcement)

Recall number: Z-2805-2026

Status: ongoing

SafeCheck provides drug interaction information from FDA-approved product labeling (openFDA). This is for informational purposes only and is not medical advice. Always consult your healthcare provider before making medication decisions.

Permobil Inc Medical Device Recall: M5 Corpus Power Wheelchair REF,120032 | SafeCheck