Medical Device Recall: Ingenuity CT CT5000 Ingenuity Plus CT5000 Ingenuity Pro CT5000 Ingenuity Premium Ingenuity CT Upgrades Ingenuity CT China Ingenuity Core 128 Elite Brazil

PHILIPS MEDICAL SYSTEMS · June 11, 2026

Reason for Recall

The clinical CT scan may not automatically trigger after the Bolus Tracking scan and may need to be triggered manually. Additional Bolus Tracking shots may occur. This may result in the clinical acquisition being inadequate for diagnosis, causing a potential for patient rescan and receiving additional radiation.

Distribution

US Nationwide Distribution including Puerto Rico

States Affected

NATIONWIDE

Quantity Affected

182

Source: FDA (openFDA Device Enforcement)

Recall number: Z-2652-2026

Status: ongoing