Medical Device Recall: Philips Spectral CT on Rails. Model Number: 728334.
PHILIPS MEDICAL SYSTEMS · March 7, 2026
Reason for Recall
Philips has identified three software issues: 1. During a continuous CT (CCT) scan, there is the potential that the Gantry could remain at the current scan position after pressing Go. 2. Potential where the message of Previous Surview Exists Select Previous Surview? <Yes> or <No> should display but does not appear. 3. After performing Surview and planning the Brain Helical acquisition by setting Brain Area DoseRight Index to increase the dose, there is the potential that the Define Head Area option in the context menu is grayed out.
Distribution
Worldwide - US Nationwide distribution in the states of Minnesota and the countries of France and Netherlands.
States Affected
NATIONWIDE
Quantity Affected
3 systems
Source: FDA (openFDA Device Enforcement)
Recall number: Z-1825-2026
Status: ongoing
SafeCheck provides drug interaction information from FDA-approved product labeling (openFDA). This is for informational purposes only and is not medical advice. Always consult your healthcare provider before making medication decisions.