Class II

Medical Device Recall: SimplyGo Mini, Standard Battery Kit; Product code: 1116816; The affected battery kits are sold as accessories for the SimplyGo Mini: SimplyGo Mini Portable Oxygen Concentrator Product code: 1113

Philips Respironics, Inc. · October 7, 2025

Reason for Recall

Certain serial numbers of the SimplyGo Mini standard battery kit contain a defective chip which can lead the battery to either fail to charge or lose voltage output when in use in the device. If the fault occurs while the device is running on battery power, the SimplyGo Mini will shut down and stop providing oxygen to the user. This may happen at a temperature within the specified operating temperature range of the SimplyGo Mini.

Product Description

SimplyGo Mini, Standard Battery Kit; Product code: 1116816; The affected battery kits are sold as accessories for the SimplyGo Mini: SimplyGo Mini Portable Oxygen Concentrator Product code: 1113600

Distribution

US Nationwide distribution in the states of GA, ID, KY, NJ, NY and the country of Canada.

States Affected

NATIONWIDE

Quantity Affected

56 units (55 US, 1 OUS)

Source: FDA (openFDA Device Enforcement)

Recall number: Z-0481-2026