Class I

Medical Device Recall: Trilogy Evo. Model Numbers: DS2110X11B, UDS2110X11B, RDS2110X11B, CA2110X12B, JP2110X16B, IN2110X15B, DE2110X13B, AU2110X15B, BL2110X15B, EU2110X15B, BR2110X18B, LA2110X15B, EE2110X15B, FR2110X14B, GB

Philips Respironics, Inc. · October 7, 2024

Reason for Recall

Potential for in-line nebulizer configuration in certain locations to result in aerosol deposit accumulation on the device's internal flow sensor, which could lead to inaccurate flow measurements.

Product Description

Trilogy Evo. Model Numbers: DS2110X11B, UDS2110X11B, RDS2110X11B, CA2110X12B, JP2110X16B, IN2110X15B, DE2110X13B, AU2110X15B, BL2110X15B, EU2110X15B, BR2110X18B, LA2110X15B, EE2110X15B, FR2110X14B, GB2110X15B, ND2110X15B, ES2110X15B, IA2110X15B, IT2110X21B, KR2110X15B, TR2110X15B, EU2110X19, FP2110X10, PP2110X10, RBRBR2110X18B. Intended for invasive and non-invasive ventilation of pediatric through adult patients weighing at least 2.5 kg in institutional, home, and non-emergency transport settings.

Distribution

Worldwide distribution - US Nationwide and the countries of AE, AF, AR, AT, AU, BD, BE, BG, BH, BO, BR, BS, CA, CH, CL, CN, CO, CZ, DE, DK, DZ, EC, EE, EG, ES, FI, FR, GB, GE, GF, GI, GR, GT, HK, HR, HU, ID, IE, IL, IN, IT, JE, JO, JP, KE, KR, KW, LB, LT, LY, MA, ME, MK, MM, MQ, MT, MX, MY, NG, NL, NO, NP, NZ, OM, PA, PE, PF, PH, PK, PL, PT, PY, QA, RE, RO, RS, RU, SA, SE, SG, SI, SK, TH, TN, TR, TW, TZ, UA, UY, VN, YT, ZA, ZW.

States Affected

NATIONWIDE

Quantity Affected

82,800 units