Medical Device Recall: Trilogy Evo O2. Model Numbers: FP2100X10, DS2100X11B, IN2100X15B, DE2100X13B, IA2100X15B, EU2100X15B, LA2100X15B, EE2100X15B, FX2100X15B, ES2100X15B, ND2100X15B, FR2100X14B, IT2100X21B, JP2100X16B, IN
Philips Respironics, Inc. · October 7, 2024
Reason for Recall
Potential for in-line nebulizer configuration in certain locations to result in aerosol deposit accumulation on the device's internal flow sensor, which could lead to inaccurate flow measurements.
Product Description
Trilogy Evo O2. Model Numbers: FP2100X10, DS2100X11B, IN2100X15B, DE2100X13B, IA2100X15B, EU2100X15B, LA2100X15B, EE2100X15B, FX2100X15B, ES2100X15B, ND2100X15B, FR2100X14B, IT2100X21B, JP2100X16B, IN2100X19, EU2100X19. Intended for invasive and non-invasive ventilation of pediatric through adult patients weighing at least 2.5 kg in institutional, home, and non-emergency transport settings.
Distribution
Worldwide distribution - US Nationwide and the countries of AE, AF, AR, AT, AU, BD, BE, BG, BH, BO, BR, BS, CA, CH, CL, CN, CO, CZ, DE, DK, DZ, EC, EE, EG, ES, FI, FR, GB, GE, GF, GI, GR, GT, HK, HR, HU, ID, IE, IL, IN, IT, JE, JO, JP, KE, KR, KW, LB, LT, LY, MA, ME, MK, MM, MQ, MT, MX, MY, NG, NL, NO, NP, NZ, OM, PA, PE, PF, PH, PK, PL, PT, PY, QA, RE, RO, RS, RU, SA, SE, SG, SI, SK, TH, TN, TR, TW, TZ, UA, UY, VN, YT, ZA, ZW.
States Affected
NATIONWIDE
Quantity Affected
24,249 units
Source: FDA (openFDA Device Enforcement)
Recall number: Z-0290-2025
Status: ongoing
SafeCheck provides drug interaction information from FDA-approved product labeling (openFDA). This is for informational purposes only and is not medical advice. Always consult your healthcare provider before making medication decisions.