Class II

Medical Device Recall: STERILIZABLE INTERNAL DEFIBRILLATION PADDLES REF LP20 STERILIZABLE INTERNAL DEFIBRILLATION PADDLES REF LP15 Defibrillation is indicated for the termination of certain potentially fatal arrhythmias

Physio-Control, Inc. · August 9, 2024

Reason for Recall

Due to customer complaints and out of box failures, the firm is conducting a voluntary recall as a result of product failures (cracks and over-molding separation of the sterilizable internal defibrillation paddles.

Product Description

STERILIZABLE INTERNAL DEFIBRILLATION PADDLES REF LP20 STERILIZABLE INTERNAL DEFIBRILLATION PADDLES REF LP15 Defibrillation is indicated for the termination of certain potentially fatal arrhythmias, such as ventricular fibrillation and symptomatic ventricular tachycardia. Synchronized cardioversion is indicated for the treatment of atrial fibrillation, atrial flutter, paroxysmal supraventricular tachycardia, supraventricular tachycardia, and in relatively stable patients, ventricular tachycardia.

Distribution

Worldwide - U.S. Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, PA, RI, SC, SD, TN, TX, VA, VT, WA, WI, and WV. The countries of Australia, Canada, Netherlands, and New Zealand.

States Affected

NATIONWIDE

Quantity Affected

3,617 devices (US-3315 and O.U.S - 302 devices)

Source: FDA (openFDA Device Enforcement)

Recall number: Z-0142-2025

Status: ongoing