Class II

Medical Device Recall: Biomet 3i Certain-compatible 4.1 mm Milled Ti Abutment REF 9001767-F Preat milled Abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arc

Preat Corp · April 2, 2025

Reason for Recall

Due to manufacturing issues, abutment products were manufactured with screw seat located too high, which may cause the mating screw to have a shorter engagement length into the associated implant.

Product Description

Biomet 3i Certain-compatible 4.1 mm Milled Ti Abutment REF 9001767-F Preat milled Abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for single-unit or multi-unit prosthetic restorations.

Distribution

U.S. Nationwide distribution in the states of AZ, CA, GA, MO, NJ, NY, OH, PA, RI, TN,TX, and WA.

States Affected

NATIONWIDE

Quantity Affected

145 abutments

Source: FDA (openFDA Device Enforcement)

Recall number: Z-1987-2025

Status: ongoing

SafeCheck provides drug interaction information from FDA-approved product labeling (openFDA). This is for informational purposes only and is not medical advice. Always consult your healthcare provider before making medication decisions.