Class II

Medical Device Recall: The Sofia 2 Flu + SARS Antigen FIA employs immunofluorescence technology in a sandwich design that is used with Sofia 2 to detect nucleocapsid protein from influenza A, influenza B, and SARS-CoV-2. Th

Quidel Corporation · January 3, 2024

Reason for Recall

One lot of the Sofia 2 Flu + SARS Antigen FIA (Lot 709083) was labelled with incorrect expiration dates. The kits contain pouched cassettes labeled with an expiration date of May 31, 2025, however the correct expiration date for the pouches is September 12, 2024. The kits are labeled with an expiration date of April 30, 2025, however, the correct expiration date for the kits is September 12, 2024. The issue was discovered on November 28, 2023 by a supply chain employee who determined that the kits had expiration dates beyond the FDA authorized shelf life.

Product Description

The Sofia 2 Flu + SARS Antigen FIA employs immunofluorescence technology in a sandwich design that is used with Sofia 2 to detect nucleocapsid protein from influenza A, influenza B, and SARS-CoV-2. This test allows for the detection of SARS-CoV and SARS-CoV-2, but does not differentiate between the two viruses.

Distribution

US Nationwide distribution in the states of FL, TN, IL, LA, NY, IN.

States Affected

NATIONWIDE

Quantity Affected

312 Kits

Source: FDA (openFDA Device Enforcement)

Recall number: Z-1086-2024

Status: ongoing

SafeCheck provides drug interaction information from FDA-approved product labeling (openFDA). This is for informational purposes only and is not medical advice. Always consult your healthcare provider before making medication decisions.