Class II
Medical Device Recall: Microalbumin Calibrator Series (mALB CAL)-IVD used to calibrate the Microalbumin Assay Catalog Number: MA1567
Randox Laboratories Ltd. · January 31, 2024
Reason for Recall
Microalbumin Calibrator Series, MA1567 & MA2426, is running with a negative bias compared to other methods and may misclassify patient results. Users may experience a shift in Quality Control and patient sample recovery of up to +12%.
Distribution
US Nationwide distribution in the states of GA, MA, WV including PR.
States Affected
NATIONWIDE
Quantity Affected
8 units
Source: FDA (openFDA Device Enforcement)
Recall number: Z-1593-2024
Status: ongoing
SafeCheck provides drug interaction information from FDA-approved product labeling (openFDA). This is for informational purposes only and is not medical advice. Always consult your healthcare provider before making medication decisions.