Class II
Medical Device Recall: Microalbumin Calibrator Series (mALB CAL)-IVD to calibrate the Microalbumin Assay Catalog Number: MA2426
Randox Laboratories Ltd. · January 31, 2024
Reason for Recall
Microalbumin Calibrator Series, MA1567 & MA2426, is running with a negative bias compared to other method may misclassify patient results. Users may experience a shift in Quality Control and patient sample recovery of up to +12%.
Distribution
US Nationwide distribution in the states of GA, MA, WV including PR.
States Affected
NATIONWIDE
Quantity Affected
6 units