Class II

Medical Device Recall: Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 17.0.0, 17.0.1 Software Version: RayStation v2025, RayStation v2025 SP1 Product Description: Radiation Therapy Treatme

RAYSEARCH LABORATORIES AB · November 28, 2025

Reason for Recall

Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.

Product Description

Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 17.0.0, 17.0.1 Software Version: RayStation v2025, RayStation v2025 SP1 Product Description: Radiation Therapy Treatment Planning System Component: No

Distribution

Refer to the attached documents: Distribution list 167168 US - updated, Distribution list 167168 OUS

Quantity Affected

0

Source: FDA (openFDA Device Enforcement)

Recall number: Z-1110-2026

Status: ongoing