Medical Device Recall: RayStation/RayPlan. Model/Catalog Numbers: 11.0.0, 11.0.1, 11.0.3, 11.0.4. Software Version: RayStation 11A, RayStation 11A SP1, RayStation 11A SP2, RayStation 11A SP3. Radiation Therapy Treatmen
RAYSEARCH LABORATORIES AB · November 28, 2025
Reason for Recall
Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.
Product Description
RayStation/RayPlan. Model/Catalog Numbers: 11.0.0, 11.0.1, 11.0.3, 11.0.4. Software Version: RayStation 11A, RayStation 11A SP1, RayStation 11A SP2, RayStation 11A SP3. Radiation Therapy Treatment Planning System
Distribution
Refer to the attached documents: Distribution list 167168 US - updated, Distribution list 167168 OUS
Quantity Affected
10 units
Source: FDA (openFDA Device Enforcement)
Recall number: Z-1103-2026