Class I

Medical Device Recall: Resmed Astral 100/150 Resmed Astral 100/50 is a ventilator system that provides continuous or intermittent ventilatory support for patients weighing more than 11lb (5kg) who require mechanical vent

ResMed Ltd. · June 25, 2026

Reason for Recall

Due to a Supercapacitor electrolyte leakage on Printed Circuit Board Assembly (PCBA) may damage the charger chip, causing fail-safe state and interruption of ventilation therapy.

Product Description

Resmed Astral 100/150 Resmed Astral 100/50 is a ventilator system that provides continuous or intermittent ventilatory support for patients weighing more than 11lb (5kg) who require mechanical ventilation.

Distribution

Worldwide - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The countries of Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bhutan, Brazil, Bulgaria, Canada, Chile, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, French Guiana, French Polynesia, Germany, Great Britain, Greece, Guadeloupe, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Islamic Republic of Iran, Israel, Italy, Japan, Jordan, Kenya, Republic of Korea, Kuwait, Latvia, Lebanon, Lithuania, Malaysia, Maldives, Malta, Martinique, Mauritius, Mexico, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Norway, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Saint Martin (French part), Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Arab Emirates, Uruguay, and Viet Nam.

States Affected

NATIONWIDE

Quantity Affected

168,069 devices globally

Source: FDA (openFDA Device Enforcement)

Recall number: Z-2735-2026

Status: ongoing