Class II

Medical Device Recall: Curaplex Field Cric Kit with ET Tube, Item Number 36611. The kit provides healthcare professionals with an assortment of supplies to perform a cricothyrotomy.

Sarnova HC, Llc · November 6, 2023

Reason for Recall

It was determined that the Endotracheal Tube Holder included in certain kits is not compatible with a cricothyrotomy procedure, leaving practitioners without the necessary means to secure the inserted endotracheal tube.

Distribution

US Nationwide

States Affected

NATIONWIDE

Quantity Affected

1063 kits

Source: FDA (openFDA Device Enforcement)

Recall number: Z-0720-2024

Status: ongoing