Class II

Medical Device Recall: Mariner RDX MIS Polyaxial Head, REF: MX1-001010, part of Mariner RDX MIS Alpha Instruments and Implants, REF: AMRDXMIS, which are pedicle screw systems.

SEASPINE ORTHOPEDICS CORPORATION · February 28, 2024

Reason for Recall

Screw heads, of pedicle screw systems, have partial thread within the screw head that pose a risk of intraoperative cross-threading within the locking cap/screw head interface.

Distribution

US: CA, FL, CO, TX

States Affected

CA, CO, FL, TX

Quantity Affected

444 Screw Heads

Source: FDA (openFDA Device Enforcement)

Recall number: Z-1620-2024