Class II

Medical Device Recall: Autosampler, Models: SIL-20ACXR CL, SIL-20ACHT CL, SIL-30AC CL

Shimadzu Corp. Analytical & Measuring Instruments Div. · October 9, 2024

Reason for Recall

There have been cases of failure of the main board due to condensation water drained from the Peltier dehumidifier, leaked out from the tray, resulting in a short circuit. In the serious case, the main board and adjacent cables are partly charred. In the minor case, the Y motor and its surrounding die cast part become rusty. These failures could result in delay in patient sample testing.

Distribution

Worldwide - US Nationwide distribution in the states of MS, AR, TX, CA, MA, MI, FL, OK, WI, IN, MN, OH, NJ, NY, IL, KY, NV, HI, SD, ID, MI, UT, WA, MO, VA, GA and the countries of DE, CN, JP, UY.

States Affected

NATIONWIDE

Quantity Affected

179

Source: FDA (openFDA Device Enforcement)

Recall number: Z-0553-2025