Class II

Medical Device Recall: Trinias unity, Digital Angiography System with Catheterization Table, Model: KS-100 This system consists of the following main components: C arm, catherization table, image processing equipment, x-ray

SHIMADZU CORPORATION MEDICAL SYSTEMS DIVISION · September 6, 2024

Reason for Recall

If digital angiography system is used with a specific catheterization table and the operator directs simultaneous movement of the table and the C-arm using direct memory, then direct memory button is released, the table may continue to move, which the operator can stop by pressing the stop button.

Product Description

Trinias unity, Digital Angiography System with Catheterization Table, Model: KS-100 This system consists of the following main components: C arm, catherization table, image processing equipment, x-ray high voltage generator, x-ray tube. There are several types of each component.

Distribution

US Nationwide distribution in the states of MA, IL, SC, KY, SC, OH, TN.

States Affected

NATIONWIDE

Quantity Affected

13

Source: FDA (openFDA Device Enforcement)

Recall number: Z-0139-2025

Status: ongoing