Class II

Medical Device Recall: ADVIA Chemistry Microalbumin_2 (¿ALB_2) -For in vitro diagnostic use in the quantitative measurement of microalbumin in human urine on ADVIA Chemistry systems/ Atellica CH and Atellica CI Analyzers.

Siemens Healthcare Diagnostics, Inc. · August 15, 2024

Reason for Recall

Atellica CH Microalbumin_2 (¿ALB_2) and the ADVIA Chemistry Microalbumin_2 (¿ALB_2) are not meeting the High-Dose Hook Effect claim/Prozone Effect claim as stated in the Instructions for Use (IFU) on the Atellica CH, Atellica CI Analyzers, ADVIA 1800 Chemistry Systems, ADVIA 2400 Chemistry Systems, and ADVIA Chemistry XPT Systems, may result in erroneously depressed microalbumin patient results.

Distribution

Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belgium, Bosnia Herzeg, Brazil, Canada, Chile, Colombia, Croatia, Cura¿ao, St. Eus, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, Hong¿Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kenya, Kuwait, Latvia, Lithuania, Malaysia, Maldives, Mexico, Morocco, Netherlands, New¿Zealand, Norway, Oman, P.R.¿China, Pakistan, Paraguay, Peru, Poland, Portugal, Qatar, Republic¿Korea, Romania, Russian¿Fed., Saudi¿Arabia, Serbia, Singapore, Slovakia, Slovenia, South¿Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, U.A.E, United¿Kingdom, Uruguay, Vietnam, Zaire. ¿¿¿¿¿¿

States Affected

NATIONWIDE

Quantity Affected

19645 units

Source: FDA (openFDA Device Enforcement)

Recall number: Z-3170-2024

Status: ongoing