Medical Device Recall: Atellica¿ CH ¿2-Microglobulin (B2M) - In vitro diagnostic use in the quantitative determination of ¿2-microglobulin in human serum and plasma. Siemens Material Number (SMN): 11097635
Siemens Healthcare Diagnostics, Inc. · October 21, 2024
Reason for Recall
Potential for falsely depressed auto-diluted results for samples above the measuring interval of 18.00 mg/L, up to 30.00 mg/L, when using the Atellica CH ¿2-Microglobulin (B2M) reagent.
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Bahrain, Belgium, Brazil, Bulgaria, Canada, Chile, Colombia, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, Germany, Greece, Hungary, India, Iran, Israel, Italy, Ivory Coast, Japan, Jordan, Kuwait, Mexico, Morocco, Oman, P.R. China, Pakistan, Paraguay, Poland, Portugal, Saudi Arabia, Saudi Arabia, Serbia, Slovakia, South Africa, Spain, Switzerland, Thailand, Turkey, U.A.E., Uganda, United Kingdom, Uruguay, Uzbekistan, Vietnam, Zaire.
States Affected
NATIONWIDE
Quantity Affected
11330 units
Source: FDA (openFDA Device Enforcement)
Recall number: Z-0599-2025
Status: ongoing