Class II

Medical Device Recall: Atellica IM CA 19-9 (250 Test - REF)-In vitro diagnostic use in the quantitative measurement of the CA 19-9 tumor-associated antigen in human serum using the Atellica CI and Atellica IM Analyzers

Siemens Healthcare Diagnostics, Inc. · July 26, 2024

Reason for Recall

Positive bias for some samples around the CA 19-9 upper limit of normal of 35 U/m as compared to the previous lot. The sample bias observed was greatest in the Asia Pacific sample population. The observed bias is not proportional across the assay measuring interval and is less pronounced for sample results above a concentration of 110 U/mL.

Product Description

Atellica IM CA 19-9 (250 Test - REF)-In vitro diagnostic use in the quantitative measurement of the CA 19-9 tumor-associated antigen in human serum using the Atellica CI and Atellica IM Analyzers SMN: 11206239

Distribution

Worldwide - US Nationwide distribution

States Affected

NATIONWIDE

Source: FDA (openFDA Device Enforcement)

Recall number: Z-3102-2024

Status: ongoing