Medical Device Recall: Atellica IM CA 19-9 (50 Test)-In vitro diagnostic use in the quantitative measurement of the CA 19-9 tumor-associated antigen in human serum using the Atellica CI and Atellica IM Analyzers SMN: 10995
Siemens Healthcare Diagnostics, Inc. · July 26, 2024
Reason for Recall
Positive bias for some samples around the CA 19-9 upper limit of normal of 35 U/m as compared to the previous lot. The sample bias observed was greatest in the Asia Pacific sample population. The observed bias is not proportional across the assay measuring interval and is less pronounced for sample results above a concentration of 110 U/mL.
Product Description
Atellica IM CA 19-9 (50 Test)-In vitro diagnostic use in the quantitative measurement of the CA 19-9 tumor-associated antigen in human serum using the Atellica CI and Atellica IM Analyzers SMN: 10995490
Distribution
Worldwide - US Nationwide distribution
States Affected
NATIONWIDE
Source: FDA (openFDA Device Enforcement)
Recall number: Z-3100-2024
Status: ongoing