Class II

Medical Device Recall: Dimension Creatinine Flex reagent cartridge. Material Number: 10872079.

Siemens Healthcare Diagnostics, Inc. · March 18, 2026

Reason for Recall

Siemens Healthcare Diagnostics is recalling certain lots of Dimension Creatinine Flex reagent cartridge (Dimension CRE2) because a potential imprecision in the Dimension CRE2 quality control (QC) and patient sample results using lot numbers GA6307 and BA7005 on the Dimension System. This imprecision may lead to erroneously decreased or increased creatinine patient results

Distribution

Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Bahamas, Brazil, Cambodia, Canada, Chile, China, Columbia, Costa Rica, Ecuador, Egypt, Germany, Honduras, India, Indonesia, Korea, Kuwait, Lebanon, Malaysia, Mexico, Nicaragua, Pakistan, Paraguay, Philippines, Qatar, Saudi Arabia, Serbia, Slovenia, South Africa, Sri Lanka, Taiwan, Thailand, U.A.E., Ukraine, Uruguay, Vietnam, Yemen.

States Affected

NATIONWIDE

Quantity Affected

18,080 units

Source: FDA (openFDA Device Enforcement)

Recall number: Z-1980-2026

Status: ongoing