Class II

Medical Device Recall: epoc BGEM BUN Test Card [25pk]. Material Number: 10736515.

Siemens Healthcare Diagnostics Inc · July 7, 2025

Reason for Recall

Siemens Healthcare Diagnostics Inc. has confirmed that a negative sodium bias may exist in some epoc test card lots. All other analytes are performing as intended. The observed average bias for sodium was -4.4 mmol/L. The maximum bias observed was -14 mmol/L, which occurred at a higher sodium concentration around 150 mmol/L. No positive bias has been observed. Quality Control (QC) fluid may or may not detect the issue depending on the bias magnitude and the frequency of QC testing.

Distribution

Worldwide distribution - US Nationwide and the countries of Argentina, Austria, Bahrain, Bolivia, Brazil, Canada, Canary Islands, Chile, Croatia, Denmark, Ecuador, Estonia, France, Fren. Polynesia, Germany, Greece, Guatemala, Hungary, Iceland, India, Indonesia, Ireland, Israel, Italy, Jordan, Kazakhstan, Lithuania, Malaysia, Martinique, Mexico, Morocco, Netherlands, New Caledonia, Paraguay, Philippines, Poland, Portugal, Qatar, Rep. Korea (S), Romania, Russian Fed., San Marino, Slovakia, Slovenia, South Africa, Spain, Srilanka, Sweden, Switzerland, Taiwan, Thailand, Turkey, Uganda, Ukraine, Unit.Arab Emir., United Kingdom, Uruguay, Japan.

States Affected

NATIONWIDE

Quantity Affected

37,554 units

Source: FDA (openFDA Device Enforcement)

Recall number: Z-2272-2025

Status: ongoing

SafeCheck provides drug interaction information from FDA-approved product labeling (openFDA). This is for informational purposes only and is not medical advice. Always consult your healthcare provider before making medication decisions.

Siemens Healthcare Diagnostics Inc Medical Device Recall: epoc BGEM BUN Test Card [25pk]. Material Number: 10736515. | SafeCheck